MICROSCOPY
Most cells are
exceedingly small in size, and all cell components are still smaller and almost
transparent to visible light. So, unaided human eye cannot resolve
them. Resolution is the ability to discriminate between two closely adjacent points
or objects as separate units, rather than viewing them as a single unit.
The lowest minimum distance between two objects, required for their
resolution (discrimination), is known as limit of resolution. The normal resolving power of human eye is 100 microns.
Microscopes
(Gk. mikros = minute, skipein= to see) are the instruments used to
produce magnified visual or photographic images of very minute objects. The
ratio between the size of the microscopically magnified image and that of the
image formed in the retina of an unaided eye is called the magnifying
power of the microscope.
The resolution
of a microscope depends directly on its magnifying and resolving
powers. If the resolving power is very low, then the images of close by
objects would overlap, making the vision blurred and indistinct. On the other hand,
a very high resolving power would make the images of adjacent
objects sharply distinct from each other.
Resolution
of a microscope
The resolving
power of a microscope depends upon the wavelength (λ) of the illuminating agent and the
numerical aperture (NA- light-gathering power) of the objective lens. Shorter
the wavelength, greater would be the numerical aperture and higher
would be the resolving power. The resolving power of a microscope is
inversely proportional to the limit of resolution of its objective lenses. So, the
resolving power can be increased by decreasing the limit of resolution i.e., by
using illuminating agents of shorter wavelength.
The resolving
power of a microscope can be determined by applying the formula
Numerical Aperture
The resolving power of a light microscope depends on the wavelength of light used and the numerical aperture (NA) of the objective lenses
The numerical aperture of a lens can be increased by
Numerical Aperture
The resolving power of a light microscope depends on the wavelength of light used and the numerical aperture (NA) of the objective lenses
The numerical aperture of a lens can be increased by
·
increasing the size of the lens opening and/or
·
increasing the refractive index of the material between
the lens and the specimen.
The larger the
numerical aperture the better the resolving power. It is important to
illuminate the specimens properly to have higher resolution. The concave mirror
in the microscope creates a narrow cone of light and has a small numerical
aperture. However, the resolution can be improved with a sub stage condenser. A
wide cone of light through the slide and into the objective lens increases the
numerical aperture there by improves the resolution of the microscope.
Based on the
source of illumination two major groups of microscopes can be recognized,
namely light microscopes and electron microscopes.
Light microscopes
These are the
microscopes in which the object is illuminated by visible light, and magnified
by an optical lens system.
Bright field microscopes
These are the
microscopes in which the object is illuminated against a bright background. Bright field microscopes are of two kinds, simple and compound.
(a) Simple
light microscope
Simple microscope consists of a
single convex glass, which can produce only
a slightly magnified visual image of the object. In a simple microscope the image appears on the
same side where the lens and the object are placed, and it cannot be
projected on a screen. So the image is
virtual. It is upright (erect), and not upside down (inverted). Simple
hand lenses, dissection microscopes, etc. are simple microscopes
(b)
Compound light microscope
Compound
microscopes are the microscopes in which there is a series of lenses, in
place of the single magnifying glass in a simple microscope. The optical system
of an ordinary light microscope has two magnifying lenses namely an
objective lens, seen close to the object, and an ocular lens or eye piece, seen
close to the observer's eye. After illuminating the object, the light rays pass
through the objective lens. The objective lens produces a slightly magnified, real
and inverted initial image of the object. This initial image serves as the
object of the ocular lens, which produces a further magnified virtual
and inverted image on the retina of the observer's eye.
Condenser lens system
Conventional light microscopes do not
have devices to improve the quality of the light beam. In modern
microscopes it is effected with the help of a condenser lens system, fixed just below the platform of the microscope. It consists of a condenser lens and a diaphragm,
placed in between the light source and the
object. Diaphragm has an aperture to
control the incident light. The lens
focuses the illuminating light on
the object. High resolving power can be obtained by modifying the
aperture of the condenser system.The resolving power of a compound microscope depends on visible light whose wavelength is much higher, ranging between 4,000 and 8,000 A0. So, compound light microscopes cannot resolve objects closer than half the wavelength
of visible light.
Oil immersion: This is a microscopic technique that
uses high powered objective lenses, called oil immersion objectives. Oil
immersion objective is a microscopic objective in which the front part of the
lens gets immersed in a liquid, when lowered. In this case a thin film of
immersion oil (such as cedar wood
oil) is placed on the cover slip, in between the objective lens and the specimen. This oil has precisely the same
refractive index as that of the glass lens,
and it increases the numerical aperture by letting a wide angle of rays to enter the objective lens. This, in turn, enhances
the resolving power and thereby gives
the maximum magnification obtainable with a light microscope.
Darkfield microscopes
Darkfield
microscopes are the light microscopes in which the object is
illuminated against a dark background by scattering light rays. They are used for observing transparent
and semitransparent objects, that are not
readily visible in a bright background. Visibility naturally depends
upon a contrast between the object and its background. In the case of transparent objects this contrast can be
increased by using a dark background. The ordinary light microscope can be
equipped for darkfield illumination
either by replacing the ordinary condenser by a darkfield condenser or by placing an opaque central disc in the centre of the ordinary condenser. This causes a cone of light to illuminate the object. The darkfield condenser illuminates the object only obliqoely. It prevents the central rays from the mirror from passing through it to illuminate the object. At the same time, it deflects the peripheral rays obliquely on to the transparent object from where they get scattered. Thus, the field of vision is dark, and not bright. The scattered rays from the object pass through the objective lens with the result that the
object becomes visible as a bright speck in a dark background. It appears that in darkfield microscopes no direct light
enters the objective lens, but only the
scattered light. Darkfield microscopes are useful in detecting very small and
transparent objects, and also objects that are poorly stained.
C. Fluorescence microscopes
These are the
microscopes whose functioning is based on the phenomenon of
fluorescence (the power to emit visible light). In them ultraviolet rays are the
illuminating agents. When they illuminate the object, fluorescence occurs in
the object. This fluorescence is of two kinds, autofluorescence and secondary fluorescence. In autofluorescence some
chemical compounds get excited at the
molecular level, when ultraviolet rays illuminate them. Consequently, they become luminous and emit visible
rays. In other words, they absorb
ultraviolet rays, and in return, emit visible light rays. This is the case with chlorophyll, porphyrin, ribofiavin, etc.
In secondary fluorescence luminosity is induced in certain substances by
staining them with fluorescent dyes, known
as fluorochromes. Consequently, they glow against a dark background, making their detection very easy. These
dyes combine with non-fluorescent
compounds, such as proteins and carbohydrates, and make them fluorescent.
Fluorescent dyes absorb light rays of a particular wavelength and emit them as light rays of another wave length,
Fluorescein (emits red light when excited with blue light), rhodamine
(emits red-green light when excited with green-yellow light), etc are
fluorescent dyes
Both in auto
and secondary fluorescence the fluorescent rays are detected by the
microscope and the chemical nature of the object is determined. Thus. fluorescence microscopes are highly
useful in testing the chemical nature of the objects and also in locating
cellular components based on their chemical nature.
An ordinary microscope can be converted to a fluorescence microscope by
introducing a special filter between the source of illumination and the
object. There are two kinds of filters, namely excitation filters and barrier filters. Excitation filters transmit only the excitation radiations to the object; all
other radiations would be
"filtered" out or prevented from passing to the object. Barrier Filters prevent the excessive flow of excitation
radiations to protect the eye of the
observer especially from ultraviolet rays).
Electron microscopes (EM)
These are the optical
instruments used for the direct study of the ultra structure of
biological systems. The first functional EM was made by Knoll and Ruska in
1928. (Ruska
shared the 1986 Nobel Prize in
Physics with Binning and Rohrer).
Electron microscopes have exceptionally high magnifying and
resolving powers. Their magnification is up
to 500,000 times, and resolution up to 2-5A. This high resolution is due to the very short
wavelength of the elections, that are
used as illuminating agents.
EM consists of
an evacuated tube in which a beam of accelerated electrons interacts with the object. The
electrons are accelerated through a
potential of 500-1500 kV in an electron gun. These electron beams can
penetrate relatively thick, hydrated and living specimens. They are deflected
and focused on to a fluorescent screen by electrostatic or electromagnetic fields, produced by electromagnetic
lenses, much in the same-way that a
beam of light is refracted by a lens in a compound microscope. An image is formed either directly or indirectly
on the fluorescent screen. In the
former, primary electrons, passing through the object, are directly focused on to the screen. But in the latter, X-rays
or secondary electrons, emitted during
the interaction of primary electrons with the object, are focused on to the screen.
EM is of two
kinds, transmission EM (TEM) and scanning EM (SEM).In TEM electrons
are transmitted right through the object to study the interior of the cell.
But, in SEM electrons are scattered or reflected back from the object to study the cell surface.
(a) Transmission electron microscope
The basic plan
of the illumination source and the lens systems of TEM closely corresponds to that of a
compound light microscope. Instead of visible light, TEM makes use of high
speed electrons of very short wavelength as
the illuminating agent. The source of these electrons is a cathode, which
acts as a thermoionic gun, called the
electron gun. It is, in fact, a metal filament placed in a vacuum tube. When it is heated, it emits streams of
electrons, which are accelerated by high voltage electrical potential This high
voltage considerably reduces their wavelength (to approximately 0.005 nm
– for visible light it is 550 nm).The
accelerated electrons tend to follow a straight path with properties similar to
those of light. From the cathode the accelerated
electrons travel through a cathode ray tube and reach the condenser lens system. The condenser system, in
turn, consists of an electromagnetic
lens, which is an electromagnetic coil, enclosed in a soft iron case.
The electromagnetic condenser focuses the
electrons on to the object. Thus, the object gets evenly illuminated by
a broad beam of electrons at 40 - 100 kV. Most of these electrons pass right through the object. But, some get
absorbed or scattered. The unscattered
electrons travel straight and reach an electromagnetic objective lens, it deflects the electron beam and produces a
direct, magnified and real image of the object (nearly 2000 times magnified). This image is further magnified by a third
electromagnetic lens, often called
the projector lens It projects the final image on a fluorescent screen for
direct viewing through an air tight "window". Often, the screen is replaced
by a photographic film with a camera attachment for recording the image. In
that case a photograph of the object is registered on the film called electron micrograph. The final image can be magnified up to
50000 times. The
camera system can further magnify it 4 - 10 times. Thus, electron micrographs
may be magnified up to 5 million times the size of the primary object.
Adjustments in focusing, and alterations in magnification, can be made by changes
in the currents of electromagnetic coils.
TEM has the
advantage of much finer resolution than light microscope But, it is not usually
suitable for living specimens, because within the specimens some electrons
are scattered and some amount of energy is lost from them. Energy
absorption by the specimen may damage it, especially at high magnification.
(b) Scanning electron microscope
SEM provides a clear surface view of
the object. In it a
very fine beam of electrons at 3 - 30 kV is made to scan a specific
surface area of the specimen, just as a screen is scanned in a television tube. The
primary electrons, emitted from it move back and forth across the specimen. During
this secondary electrons are emitted either from the surface
or from near surface of the specimen. They are collected by a photo multiplier
tube or a positively charged grid. From there the signal is transferred to a
television tube with the result that an image is displayed on the screen.
The image
produced by SEM resembles those seen in a hand lens. But, it has much finer
resolution and can be further magnified nearly 100,000 times.
Phase contrast microscopes
These are the microscopes used for
examining the components of unstained
living cells. They are designed on the principle that small phase changes or phase differences in the light rays passing
through an object can be transformed
to differences in brightness or light intensity. The phase contrast principle was first worked out by Zernike in 1940.
Living cells
are not usually coloured, but almost transparent to visible light. This
means that they are pure phase objects and the light passing through them
practically suffers no loss of intensity due to absorption. However, in reality,
the light passing through a living cell does encounter regions of different refractive
index and thickness, which alter
its velocity and direction. This difference in refractive index produces a difference in the optical path of
the light transmitted by the
different regions. As a result, the
light transmitted by the region of higher thickness and greater refractive index gets retarded in velocity and emerges out of phase in relation to the light transmitted by the region of lower thickness and low refractive index. When the difference in refractive index is small, the magnitude of the phase
change would also be small, and is
measured in wavelength ( ). Phase contrast microscope
can transform such phase changes
into corresponding variations of brightness or intensity. This, in turn,
enhances the contrast between the cell, its content and its surroundings, enabling the study of the cell in the living state.
The optical system of a phase contrast
microscope differs from that of an ordinary light microscope only in the
addition of (i) a sub stage annular diaphragm to illuminate the object with a
narrow cone of light, and (ii) a diffraction plate (phase retardation plate or
phase plate) in the objective. For enhancing the phase difference between
direct and diffracted rays phase contrast microscope has special optical
devices. To change the relative phase the diffracted waves have to be first
separated from direct waves. This is achieved by inserting a sub stage annular
diaphragm in the condenser system. This diaphragm permits the light to pass
through the condenser in the form of a hollow cone of rays, the "central
rays" being absorbed. This light cone illuminates the object. The objective lens will
focus the annulus on its upper (back) focal
plane. The direct rays passing through the thinner regions of the object will form the image of the
annulus, while the diffracted rays
passing through its denser region will spread over the entire focal
plain producing a number of overlapping images of the diaphragm.
Thus, direct and diffracted rays are spatially separated from one another. By
inserting phase
retardation plate at the back focal
plane of the objective lens the phase change
of the diffracted rays can be further increased.The phase retardation plate is
a channeled or grooved and transparent glass disc. Its channel is coated with a
material that can absorb light, but cannot
retard it. The remainder of the plate is coated with a thin film of light retarding material (such as magnesium fluoride).
The channel is so designed that it
exactly matches the image of the condenser annulus. So, the light passing directly through the specimen and the
channel will not be further retarded. On the other hand, the diffracted light
passing through the remainder of the plate gets retarded again. Thus,
the phase contrast between direct and
deviated rays gets exaggerated.The
magnification by a phase contrast microscope is not much higher than that of a compound microscope.
Still, it is much significant in the following respects :
1. It enables to
distinguish living cells from the surrounding medium
2. It is useful to
discriminate between various cell components and also to observe the
effects of chemical and physical agents on living cells.
3. All kinds of
cellular movements, including the chromosomal movements during cell
division, can be detected by phase contrast microscopy
4.
The
changes undergone by cytoskeleton during cellular movements can be observed with the help of phase
contrast microscopy.
5.
The Phase contrast microscopy
is particularly useful in observing living cells and tissues in tissue culture.
Phase contrast microscope has two
major disadvantages also: 1) It is ideal only for observing individual cells or
thin cell layers 2) It often results in halo formation, since the separation of
direct and deviated rays is only partial or incomplete.
Centrifugation
Centrifugation
Centrifugation
is the separation and sedimentation of sub- cellular fractions of the
homogenate according to their size, mass, density and specific gravity. The
instrument used for this is called centrifuge . It consists of a pivoted
cylinder, called rotor, in which the samples are placed. The rotor rotates or
spins around a central axis at very high speeds. This produces gravitational
fields, which are higher than earth's gravity. The high rotational speed of the rotor produces a
centrifugal force, directed away from the centre. It may be as much as 100,000
times greater than gravitational force. The magnitude
of the centrifugal force is expressed in terms of earth's gravitational force, namely 'g' (g = 980 cms/sec2).
The relative centrifugal force (RCF)
is directly proportional to the number of revolutions per minute (rpm-ω),
radius of rotation (r) and the gravitational force
Centrifugal force causes the sample particles to move away from the centre
and also to settle in different layers according their sedimentation ratio. Heavier particles move away first and get
deposited on the outer wall of the tube. Lighter particles follow them
in regular sequence in the order of decreasing mass.
The intensity
of the centrifugal force in relation to the sedimentation of particles can
be represented by the equation.
The rate of sedimentation of particles
in centrifugation is called sedimentation coefficient. The unit of its
measurement is called Svedberg unit(S). The suspension, left above the sediment after centrifugations,
is called supernatant. It may be subjected to repeated centrifugation, with
progressively increasing speed. As a
result, different particles settle down differently at different times, at different rates and at
different sites. This may be called differential centrifugation.
In classical
fractionation and centrifugation cell components separate into four successive
fractions in the order nuclear fraction, mitochondrial fraction, microsomal
fraction and soluble fraction. For the separation of DNA and RNAs an improved
method of centrifugation has been deviced. It is called density
gradient centrifugation.
Kinds of
centrifuges
Three major
kinds of centrifuges can be recognized, namely low-speed centrifuges, high-speed centrifuges,
and ultracentrifuges. Low-speed centrifuges
are the most ordinary types of centrifuges, used for the routine sedimentation of heavy particles. Their rotor has
a maximum speed up to 4,000 or 5,000 rpm, with RCF values up to 3000g.
Low-speed centrifuges are used at room temperature, and they have no means for
temperature control of the samples. High-speed centrifuges are advanced types of
centrifuges, with higher speed and
temperature control of the rotor chamber. They are especially important in the centrifugation of
temperature-sensitive biological samples.
Their rotor revolves at an average speed of 20,000 rpm, with an RCF value up to 50000g.
Ultracentrifuges are the most
sophisticated and refrigerated types of centrifuges.
In them the rotor rotates at extremely high speeds (75,000 rpm), producing a gravitational pull of above 50,000g.
Ultracentrifuges are used for the separation of viruses and extremely
minute sub-cellular components and also for the determination of molecular
weights. They are used for preparative
work and analytical measurements. So, there are two kinds of them, namely preparative Ultracentrifuges and
analytical Ultracentrifuges. Analytical
Ultracentrifuges are provided with window and optical system. So, sedimentation
of particles can be observed and their sedimentation rates can be measured during the run.
Methods
of centrifugation
There are two
main kinds of centrifugation, namely differential centrifugation and
density -gradient centrifugation.
(i)
Differential centrifugation
Differential
centrifugation is the successive centrifugation of the homogenate at progressively
increasing rotor speeds, leading to the successive separation of cell components. It involves the sedimentation of
particles in a medium of homogeneous
density.
(ii) Density-gradient centrifugation
Density-gradient
centrifugation is the technique used for the separation of multi component mixtures of
macromolecules and also for the measurement of sedimentation coefficients (S
values). In this case, the suspending liquid
and the multi component sample exhibit density difference. Density will be highest at the bottom of the centrifuge
tube, and lowest at the top. In between
these two extremes there will be a gradient (graded series) of density- different zones. This may be called density
gradient. The multi component sample is layered at the top of the
gradient. Ultracentrifugation at a constant
speed results in the sedimentation of the sample components according to their density. When a fraction
reaches a liquid zone, whose density
is the same as that of the fraction, the fraction stays there without any further sedimentation. In this manner different
fractions of the sample separate
into distinct density- different zones or bands.
Density-gradient
centrifugation is widely used in the separation and purification of biological
macromolecules and also for the determination of S values.
AUTORADIOGRAPHY
This is the
technique used to observe the distribution of a particular chemical substance in living cells and
tissues, and to track the routes and conversions of macromolecules in different
biochemical reactions. In autoradiography, radioisotopes
(isotopes which can emit ionizing radiations) of some elements (e.g. I4C, 3H, 32P,
35S, etc.) are introduced to the cell. Each of them gets incorporated with a specific substance and makes it
radioactive. This is called labelling of the substance with
radioisotopes, or simply radioactive labelling. Then, the movement and distribution of the labelled substance is
observed. This helps to detect the presence of the substance and also to track
the synthesis of a compound out of
it. The radioisotopes can emit one or more of the three radiations, namely alpha and beta particles and gamma rays. Most
of the isotopes used in autoradiography are beta-emitters.The cells or tissues, containing radioactively
labelled molecules, are placed in
contact with a photographic emulsion for a certain period. The ionizing radiations, emitted by the radioisotopes, blacken
the emulsion and produces an image, known as autoradiograph. A comparison of
this image with the normal cells observed under microscope will enable to
detect the location of the
radioactive isotope and also to track the route of the radioactively labelled molecule. Autoradiographs are usually
observed with the help of phase contrast
microscopes
ELECTROPHORESIS
Electrophoresis
is the analytical technique applied for the
separation of charged molecules, based on
their movement in an electric field ; positively charged particles move to the cathode, and negatively charged ones to the
anode. This makes their separation
easy. It is a very convenient technique for analysing and purifying several kinds of biomolecules, especially
peptides, proteins, nucleotides and nucleic
acids. Separation of the constituents of mixtures or solutions occurs at pH values .above or below the isoelectric point.
Electrophoresis
relies on the principle that biological molecules in solution carry a net electric charge
(except at the isoelectric point). If two electrodes are placed in such a solution, and an electric field is applied,
the charged particles move towards
oppositely charged poles at different rates and get localized in different zones. Their movement is
greatly influenced by their shape, size, molecular weight and electric
charge, and also by the matrix of the gel support and the applied electric
field. The greater the charge and the lesser
the molecular weight, the greater would the electrophoretic mobility of the particles. The charge, in turn, depends on
the pH of the buffer solution. A
molecule with a double charge will move at twice the speed of a molecule with a single charge. This difference is the basis
for the separation of the different molecules of mixtures or solutions.
It is very unlikely that two different kinds
of molecules of a mixture will have exactly the same size, molecular weight and
charge level.
Modern electrophoretic
techniques use wetted filter paper or a polymerized gel - like matrix
(acrylamide gel, starch gel, cellulose acetate gel, agarose gel, silica gel,
etc.) as an inert support medium. The sample to be analysed is applied to this
medium as a spot or as a thin band.
The movement of charged molecules in
an applied electric field is represented by
the equation
Thus , charged
particles move at a velocity which depends directly on the applied electric field or
voltage (E) and charge (q) , but inversely on a counteracting force generated by the viscous drag (f)
Types of electrophoresis
All
electrophoretic methods are based on the same principle. So, they differ from
each other only in the nature of the support medium. Some common types of
electrophoretic methods are paper electrophoresis, polyacrylamide gel
electrophoresis (PAGE - significant in the separation of proteins ), agarose
gel electrophoresis (AGE - important in the separation of nucleic acids), sodium dodecyl
sulphate - polyacrylamide gel electrophoresis
(SDS- PAGE), significant in the
measurement of the molecular weight of biomolecules), pulsed field
gel electrophoresis (PFGE), significant in the separation of large chromosomal DNA, capillary electrophoresis (CE-
combination of the resolving power of electrophoresis with the speed of
high performance liquid chromatography to
analyse very small samples ), etc.
Paper
electrophoresis
This is the electrophoretic method in which a filter paper wetted with a buffer solution forms the support medium. The paper spans between two reservoirs of the buffer
solution, with its ends
immersed in the solution. A small
amount of the solution or mixture to
be analysed is placed somewhere near the centre of the paper. Now, electrodes
are placed in the reservoirs of the
buffer and a voltage is applied. The charged constituents of the mixture or
solution move to their respective poles at different velocity. This brings about their separation.
Polyacrylamide gel electrophoresis (PAGE)
PAGE is the
electrophoretic method in which polyacrylamide gel is used as the support
medium. Polyacrylamide gel is a synthetic polymer (prepared by the free radical
polymerisation of acrylamide and the cross linking agent
methylene-bis-acrylamide). Polyacrylamide gels are produced either as columns or as
slabs. Slab gels are now more widely used than column gels. A slab gel is more
convenient, because several samples can be analysed on it; only a single sample
can be analysed on a tube (column) gel.
Acrylamide gel
is something like a spongy network. Macromolecules cannot freely
pass though it, but can only squeeze through its narrow channels. The rate of
this movement depends largely on the size and molecular weight of the
migrating molecules. Smaller the molecules and lower their molecular weight, faster
would be their movement.
In PAGE gel
column or gel slab is inserted vertically between two buffer reservoirs. The
upper reservoir usually contains the anode, and the lower one contains the
cathode. The sample to be analysed is placed on the gel slab or gel
tube and a voltage is applied. The charged molecules of the sample move to
their respective poles.PAGE is significant in that in it separation of the
sample components involves both molecular sieving and electrophoretic
movement So, it results in enhanced resolution of sample components The
order of molecular movement in gel filtration and PAGE is different. In gel
filtration large molecules move through the matrix faster than small
molecules, but the reverse is the order in PAGE. Usually, PAGE is used for the
separation of large molecules, such as proteins.
No comments:
Post a Comment
Note: Only a member of this blog may post a comment.